A circular, shear-driven pumping system combined with Fourier Transform detection has been developed for the application of chip based cyclic chromatography. Using this system, it is possible to perform an injection of a sample plug into a circular micro-channel and then drive the sample through the channel, using shear flow induced by a rotation stage. High pumping rates at uniform speeds are attainable with this system with very little heat production. Typical pumping rates of up to 1.423 mm s<sup>â 1</sup> ± 2 µm s<sup>â 1</sup> were used in experiments although much higher rates >14 mm s<sup>â 1</sup> are also possible with this system. Fluorescence detection was used to detect a sample plug of Coumarin dyes, flowing around the channel. A coating of porous polymethacrylate was used to immobilise RP-HPLC beads onto the glass surface and was applied to the glass micro-channel by selective photopolymerisation. This coating acted as a stationary phase and differences in retention time were observed for an injection of Coumarin dyes for different methanolâ water, mobile phase ratios. Full sample retention occurred for 30 : 70 (v/v) methanolâ water whereas no retention occurred for 92 : 8 (v/v) methanolâ water which can be expected for such a reverse phase, open tubular system. Fourier transform detection applied to the fully retained and non-retained cases showed frequency domain data from a single detection point corresponding to that which may be expected from shear flow theory. |
Формат |
application.pdf |
Издатель |
Royal Society of Chemistry |
Название |
Shear-driven pumping and Fourier transform detection for on chip circular chromatography applications |
Тип |
research-article |
DOI |
10.1039/b502121a |
Electronic ISSN |
1473-0189 |
Print ISSN |
1473-0197 |
Журнал |
Lab on a Chip |
Том |
5 |
Первая страница |
764 |
Последняя страница |
771 |
Аффилиация |
Yang Xin; Imperial College London, Department of Chemistry |
Аффилиация |
Jenkins Gareth; ISASâ Institute for Analytical Sciences |
Аффилиация |
Franzke Joachim; ISASâ Institute for Analytical Sciences |
Аффилиация |
Manz Andreas; ISASâ Institute for Analytical Sciences |
Выпуск |
7 |
Библиографическая ссылка |
Vilkner, Anal. Chem., 2004, 76, 3373 |
Библиографическая ссылка |
Debesset, Lab Chip, 2004, 4, 396 |
Библиографическая ссылка |
Burggraf, Sens. Actuators B, 1994, 20, 103 |
Библиографическая ссылка |
Manz, Fresenius J. Anal. Chem., 2001, 371, 195 |
Библиографическая ссылка |
Eijkel, Electrophoresis, 2004, 25, 243 |
Библиографическая ссылка |
Burggraf, J. High-Res. Chromatogr., 1993, 16, 594 |
Библиографическая ссылка |
Chen, J. Anal. Chem., 2005, 77, 658 |
Библиографическая ссылка |
Eijkel, Sens. Actuators B, 2003, 92, 215 |
Библиографическая ссылка |
Studer, Analyst, 2004, 129, 944 |
Библиографическая ссылка |
Desmet, J. Chromatogr. A, 1999, 855, 57 |
Библиографическая ссылка |
Desmet, Anal. Chem., 2000, 72, 2160 |
Библиографическая ссылка |
Crabtree, Anal. Chem., 1999, 71, 2130 |
Библиографическая ссылка |
Kaneta, Anal. Chem., 2001, 73, 540 |
Библиографическая ссылка |
Eijkel, Lab Chip, 2001, 1, 122 |
Библиографическая ссылка |
Kwok, J. Chromatogr. A, 2001, 924, 177 |
Библиографическая ссылка |
Kwok, Electrophoresis, 2001, 22, 222 |
Библиографическая ссылка |
Kwok, Anal. Chem., 2001, 73, 1748 |
Библиографическая ссылка |
Desmet, J. Chromatogr. A, 2001, 35, 111 |
Библиографическая ссылка |
Terabe, Anal. Chem., 1984, 56, 111 |