On-site monitoring of biogenic emissions from Eucalyptus dunnii leaves using membrane extraction with sorbent interface combined with a portable gas chromatograph system
Liu, Xinyu; Pawliszyn, Richard; Wang, Limei; Pawliszyn, Janusz; Liu Xinyu; Department of Chemistry, University of Waterloo; Pawliszyn Richard; Department of Chemistry, University of Waterloo; Wang Limei; Department of Chemistry, University of Waterloo; Pawliszyn Janusz; Department of Chemistry, University of Waterloo
Журнал:
Analyst
Дата:
2004
Аннотация:
Membrane extraction with sorbent interface, combined with a portable gas chromatograph system (MESI-Portable GC) for continuous on-line monitoring of biogenic volatile organic compounds (BVOCs) emissions (from leaves of Eucalytus dunnii in a greenhouse), is presented herein. A sampling chamber was designed to facilitate the extraction and identification of the BVOCs emitted by the Eucalytus dunnii leaves. Preliminary experiments, including; enrichment times, microtrap temperatures, stripping gas flow rates, and desorption temperatures were investigated to optimize experimental parameters. The main components of BVOCs released by the Eucalytus dunnii leaves were identified by comparing the retention times of peaks with those of authentic standard solutions. They were then confirmed with solid phase microextraction coupled with gas chromatography and mass spectrometry (SPME-GC-MS). BVOC emission profiles of α-pinene, eucalyptol, and γ-terpinene emitted by intact and damaged Eucalytus dunnii leaves were obtained. The findings suggest that the MESI-Portable GC system is a simple and useful tool for field monitoring changes in plant emissions as a function of time.
326.3Кб