Автор |
A A Mills |
Автор |
D W Sears |
Автор |
R Hearsey |
Дата выпуска |
1977-01-01 |
dc.description |
Apparatus for the measurement of thermoluminescence is described which generates glow curves with an overall reproducibility of +or-5%. Particular attention has been given to the design of a controller capable of accurately reproducing a chosen linear heating rate (0.1-99.9 degrees C s<sup>-1</sup>) up to a predetermined final temperature (ambient to 990 degrees C). A manually operated 'hold' facility for intermediate temperatures is also provided. The construction of the light-measuring system and checks on its stability, sensitivity and linear response to very weak sources are also described. The apparatus was designed for the study of linear samples and meteorites. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Apparatus for the measurement of thermoluminescence |
Тип |
paper |
DOI |
10.1088/0022-3735/10/1/013 |
Print ISSN |
0022-3735 |
Журнал |
Journal of Physics E: Scientific Instruments |
Том |
10 |
Первая страница |
51 |
Последняя страница |
56 |
Аффилиация |
A A Mills; Dept. of Geology & Astron., Univ. of Leicester, Leicester, UK |
Аффилиация |
D W Sears; Dept. of Geology & Astron., Univ. of Leicester, Leicester, UK |
Аффилиация |
R Hearsey; Dept. of Geology & Astron., Univ. of Leicester, Leicester, UK |
Выпуск |
1 |