dc.description |
The position sensitivity of hot-film shear probes was studied with the use of pulse-echo ultrasound. The ultrasonic probe was placed opposite the hot-film probe across a rigid tube. The accuracy of measuring hot-film probe position by ultrasonic methods was studied with the aid of a mechanical micrometer. A similar study was conducted in vivo on the thoracic aorta of dogs to maintain flush conditions between the hot-film probe and arterial wall. The sensitivity of hot-film probes to position with respect to a rigid wall was studied. Probe position was determined by ultrasonic methods. The results showed a 20% deviation in hot-film anemometer bridge voltage with a change in probe position of 0.3 mm from the wall. This could result in a 50% error in computed wall shear stress. Resolution of the method of ultrasonic positioning of the hot-film probe was determined to be 0.1 mm. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Position sensitivity of hot-film shear probes |
Тип |
paper |
DOI |
10.1088/0022-3735/13/9/018 |
Print ISSN |
0022-3735 |
Журнал |
Journal of Physics E: Scientific Instruments |
Том |
13 |
Первая страница |
973 |
Последняя страница |
976 |
Аффилиация |
G E Miller; Texas A&M Univ., College Station, TX, USA |