Автор |
C J Studman |
Автор |
S E Compton |
Дата выпуска |
1983-03-01 |
dc.description |
A low-cost turbine flowmeter has been designed for studies of water thermosyphoning. It was built from an ordinary medical 'spinhaler' inhalent impeller and was capable of measuring flows down to 1.7 ml s<sup>-1</sup>. It caused a pressure loss of 5 mm head of water (50 Pa) at a flow of 20 ml s<sup>-1</sup>, but the smallest detectable flow was then 10 ml s<sup>-1</sup>. The device could be manufactured in a small workshop without difficulty. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
A low-cost sensitive flowmeter |
Тип |
paper |
DOI |
10.1088/0022-3735/16/3/002 |
Print ISSN |
0022-3735 |
Журнал |
Journal of Physics E: Scientific Instruments |
Том |
16 |
Первая страница |
190 |
Последняя страница |
192 |
Аффилиация |
C J Studman; Agricultural Engng. Dept., Massey Univ., Palmerston North, New Zealand |
Аффилиация |
S E Compton; Agricultural Engng. Dept., Massey Univ., Palmerston North, New Zealand |
Выпуск |
3 |