Автор G H C Freeman
Автор W H King
Дата выпуска 1983-03-01
dc.description Six Cu II lines and one of Cu I from a liquid nitrogen cooled hollow cathode lamp have been studied at high resolution (+or-3 fm). The six Cu II lines can be produced with a consistency of +or-10 fm under the following conditions: hollow cathode diameter 10 mm+or-1 mm, helium carrier gas at 400 Pa+or-50 Pa, discharge current of about 200 mA. These six lines are different types of transition and so many other Cu II lines in the vacuum ultraviolet should be obtainable with equally good consistency. The behaviour of these lines under different source conditions is discussed.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Copper wavelength standards good to 10 fm in the vacuum ultraviolet
Тип paper
DOI 10.1088/0022-3735/16/3/011
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 16
Первая страница 211
Последняя страница 216
Аффилиация G H C Freeman; Div. of Quantum Metrology, NPL, Teddington, UK
Аффилиация W H King; Div. of Quantum Metrology, NPL, Teddington, UK