Автор D K Gibson
Автор I D Reid
Дата выпуска 1984-12-01
dc.description An electron energy analyser has been designed using the concept of the 'fountain spectrometer' and the 30 degrees plane electrostatic field configuration. The advantages of the design are that measurements at a number of scattering angles from 0 to 120 degrees can be made simultaneously. The performance of the analyser is illustrated by measurements of the electron distributions from H<sup>+</sup> on He collisions, which show the processes of charge exchange into the continuum and autoionisation of the doubly excited He atom.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A modified fountain spectrometer for measuring double differential cross sections in ion-atom collisions
Тип paper
DOI 10.1088/0022-3735/17/12/032
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 17
Первая страница 1227
Последняя страница 1230
Аффилиация D K Gibson; Australian Energy Comm. Res. Establ., Lucas Heights Res. Labs., Sutherland, NSW, Australia
Аффилиация I D Reid; Australian Energy Comm. Res. Establ., Lucas Heights Res. Labs., Sutherland, NSW, Australia
Выпуск 12