Автор |
H A Hoff |
Автор |
S J Rothman |
Автор |
J N Mundy |
Дата выпуска |
1985-03-01 |
dc.description |
The calibration of an optical pyrometer against a tungsten strip lamp has been extrapolated from the maximum temperature of the lamp, 2300 degrees C, to 3100 degrees C. The accuracy of the extrapolation has been shown to be within 0.65% by the use of sectored discs and a unique tungsten black-body source. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Calibration of an optical pyrometer above 2300 degrees C |
Тип |
paper |
DOI |
10.1088/0022-3735/18/3/006 |
Print ISSN |
0022-3735 |
Журнал |
Journal of Physics E: Scientific Instruments |
Том |
18 |
Первая страница |
197 |
Последняя страница |
198 |
Аффилиация |
H A Hoff; Mater. Sci. & Technol. Div., Argonne Nat. Lab., IL, USA |
Аффилиация |
S J Rothman; Mater. Sci. & Technol. Div., Argonne Nat. Lab., IL, USA |
Аффилиация |
J N Mundy; Mater. Sci. & Technol. Div., Argonne Nat. Lab., IL, USA |
Выпуск |
3 |