Дата выпуска 1986-09-01
dc.description A katharometer for determining the concentration of helium in air in the range 0-100% was coupled to an electronic feedback bridge supply circuit. It was found that an optimal feedback gain exists for which the measurement non-linearity reaches a minimum value of 1%, which is 15 times better than with the conventional constant voltage bridge supply.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A katharometer bridge circuit with linearising feedback
Тип paper
DOI 10.1088/0022-3735/19/9/008
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 19
Первая страница 695
Последняя страница 696
Аффилиация W P J Holland; Central Res. Lab., Med. Fac., Erasmus Univ., Rotterdam, Netherlands
Аффилиация W Boender; Central Res. Lab., Med. Fac., Erasmus Univ., Rotterdam, Netherlands