Автор C W van der Wal
Автор L C E Struik
Дата выпуска 1969-02-01
dc.description A simple circuit based on a Wheatstone bridge is described which enables temperature readings to be obtained directly in degrees Celsius with a platinum resistance thermometer, without the use of correction or calibration tables. Irrespective of the inaccuracy of the thermometer itself, reading accuracy is to better than ±0·1 degC for a temperature range from 0 to +500°C. Variation in the sensitivity of the bridge can be less than 15% in the same temperature range.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A direct-reading bridge for a platinum resistance thermometer
Тип paper
DOI 10.1088/0022-3735/2/2/304
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 2
Первая страница 143
Последняя страница 145
Аффилиация C W van der Wal; Centraal Laboratorium TNO, Delft, The Netherlands
Аффилиация L C E Struik; Centraal Laboratorium TNO, Delft, The Netherlands
Выпуск 2