Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор V T Morgan
Автор D E Brown
Дата выпуска 1969-09-01
dc.description Although three-terminal capacitors have been used to measure accurately small displacements in air and vacuum, it is usually impractical to use them in dielectric liquids, due to the strong dependence of the permittivity on purity and temperature. A differential-capacitance transducer in the form of a guarded disk between parallel plates was only slightly affected by variation in temperature over the range 18 to 90°C, and was unaffected by variation in the relative permittivity of air or liquid dielectric from 1 to 49. Displacements up to 2·5 mm could be measured with a precision better than 0·01 μm.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A differential-capacitance transducer for measuring small displacements
Тип paper
DOI 10.1088/0022-3735/2/9/309
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 2
Первая страница 793
Последняя страница 795
Аффилиация V T Morgan; Division of Applied Physics, National Standards Laboratory, CSIRO, Sydney, Australia
Аффилиация D E Brown; Division of Applied Physics, National Standards Laboratory, CSIRO, Sydney, Australia
Выпуск 9

Скрыть метаданые