Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор M K Khan
Автор K A MacKenzie
Автор H H Bruun
Дата выпуска 1987-08-01
dc.description Describes the results of a study into the effect of a dedicated calibration facility on the calibration relationships for hot-wire probes. Calibrations were carried out with plated and unplated normal hot-wire probes using both parallel and perpendicular support orientation in a calibration facility with variable nozzle exit area and in a wind tunnel. The results obtained clearly demonstrate that the cross sectional area of the calibration flow has a major effect on the evaluated calibration relationship. In particular the results for the parallel support orientation show a continuous variation in the interpreted velocity with cross sectional area. The difference between the 12 mm<sup>2</sup> and the 120 mm<sup>2</sup> nozzles was about 3-4% and a further difference of about 3-4% was observed for the wind tunnel comparison.
Формат application.pdf
Издатель Institute of Physics Publishing
Название The effects of blockage correction in hot-wire probe calibration facilities
Тип paper
DOI 10.1088/0022-3735/20/8/015
Print ISSN 0022-3735
Журнал Journal of Physics E: Scientific Instruments
Том 20
Первая страница 1031
Последняя страница 1035
Аффилиация M K Khan; Postgraduate Sch. of Mech. & Manuf. Syst. Eng., Bradford Univ., UK
Аффилиация K A MacKenzie; Postgraduate Sch. of Mech. & Manuf. Syst. Eng., Bradford Univ., UK
Аффилиация H H Bruun; Postgraduate Sch. of Mech. & Manuf. Syst. Eng., Bradford Univ., UK
Выпуск 8

Скрыть метаданые