An Intercomparison Between a Primary Standard Mercury Barometer and a Gas-operated Pressure Balance Standard
G N Peggs; K W T Elliott; Sylvia Lewis; G N Peggs; Division of Mechanical and Optical Metrology, National Physical Laboratory, Teddington, Middlesex, UK; K W T Elliott; Division of Mechanical and Optical Metrology, National Physical Laboratory, Teddington, Middlesex, UK; Sylvia Lewis; Division of Mechanical and Optical Metrology, National Physical Laboratory, Teddington, Middlesex, UK
Журнал:
Metrologia
Дата:
1979-04-01
Аннотация:
The purpose of this paper is to discuss the design and performance of a primary standard barometer and a primary standard pressure balance, and to describe in detail a direct intercomparison between the two instruments within their optimum operating range. The uncertainties associated with each pressure standard are discussed together with an estimate of the uncertainties associated with the intercomparison experiment itself. At a pressure of 100 kPa the two methods agree to within 1.2 ± 1.4 Pa at the 99.7% confidence level.