Автор Z Gengru
Автор H Yifei
Автор W Lu
Автор Z Guifen
Автор H D Jensen
Автор L C A Henderson
Автор R P Thompson
Дата выпуска 1997-10-01
dc.description The Josephson array voltage standard at the Beijing Institute of Radio Metrology and Measurement (BIRMM) has been used in China to realize the volt since 1992. In order to verify the accuracy and reliability of the voltage standard at the BIRMM, a comparison of the Josephson array voltage standards at 1,018 V and 10 V between the BIRMM, the Danish Institute of Fundamental Metrology (DFM) and the National Physical Laboratory (NPL), UK, has been performed. Two Zener reference standards were used to transfer the voltage between the three laboratories. The results show a difference of less than 2 × 10<sup>-8</sup>.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Comparison of Josephson array voltage standards using Zener reference standards
Тип note
DOI 10.1088/0026-1394/34/5/7
Electronic ISSN 1681-7575
Print ISSN 0026-1394
Журнал Metrologia
Том 34
Первая страница 423
Последняя страница 425
Аффилиация Z Gengru; BIRMM, DFM, NPL