Автор B K A Martensson
Дата выпуска 1969-01-01
dc.description The standard deviation of the signals recorded from thermoluminescent LiF as a function of the thermal treatment before irradiation (pre-annealing) has been investigated (see preceding abstr). For pre-annealing temperatures higher than -50 degrees C, the standard deviation increases with pre-annealing temperature and time. A consequence is that in pre-annealing, high temperatures should be used for periods as short as possible. If the thermal treatment during the readout (a few seconds at approximately 300 degrees C) is the only high temperature treatment of the dosimeter between repeated irradiations, a standard deviation of 0.22% in the measured signals can be obtained. Different factors contributing to the standard deviation are analysed and, in consequence, a sensitive method for quantitative comparison between different readout instruments is given.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Thermoluminescence of LiF: A statistical analysis of the influence of pre-annealing on the precision of measurement
Тип paper
DOI 10.1088/0031-9155/14/1/007
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 14
Первая страница 119
Последняя страница 130
Аффилиация B K A Martensson; Radiation Physics Dept., Univ. Lund, Sweden
Выпуск 1