Автор |
D E Pegg |
Автор |
A R Hayes |
Дата выпуска |
1970-07-01 |
dc.description |
A platinum resistance thermometer is described capable of reading temperatures in the range +37 degrees C to -200 degrees C with an accuracy of +or-0.1 degrees C. The sensor is a standard industrial platinum element, and the measuring bridge is a simple d.c. circuit employing a robust null-point detector. A calibration procedure involving the use of four convenient fixed points is described, and a simple method of interpolation is given. The performance of the instrument during 12 months of continuous laboratory use is described. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
A platinum resistance thermometer for use in cryobiology |
Тип |
paper |
DOI |
10.1088/0031-9155/15/3/001 |
Electronic ISSN |
1361-6560 |
Print ISSN |
0031-9155 |
Журнал |
Physics in Medicine and Biology |
Том |
15 |
Первая страница |
409 |
Последняя страница |
416 |
Аффилиация |
D E Pegg; Nat. Inst. Medical Res., London, UK |
Аффилиация |
A R Hayes; Nat. Inst. Medical Res., London, UK |
Выпуск |
3 |