Автор D E Pegg
Автор A R Hayes
Дата выпуска 1970-07-01
dc.description A platinum resistance thermometer is described capable of reading temperatures in the range +37 degrees C to -200 degrees C with an accuracy of +or-0.1 degrees C. The sensor is a standard industrial platinum element, and the measuring bridge is a simple d.c. circuit employing a robust null-point detector. A calibration procedure involving the use of four convenient fixed points is described, and a simple method of interpolation is given. The performance of the instrument during 12 months of continuous laboratory use is described.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A platinum resistance thermometer for use in cryobiology
Тип paper
DOI 10.1088/0031-9155/15/3/001
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 15
Первая страница 409
Последняя страница 416
Аффилиация D E Pegg; Nat. Inst. Medical Res., London, UK
Аффилиация A R Hayes; Nat. Inst. Medical Res., London, UK
Выпуск 3