Автор |
D O Bottrill |
Автор |
M A Jamal |
Автор |
K I Ahmed |
Дата выпуска |
1972-11-01 |
dc.description |
The circuit described has been installed in a medical linear accelerator to operate the high voltage circuit interlock whenever arcing occurs in the high voltage pulse forming network (PFN) or should free-running occur in the trigger pulse generator for the main thyratron. The PFN charging pulses are monitored by a 10 Omega resistor placed in series with the high voltage power supply capacitor and fed directly into the circuit. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
A trip-out circuit for arcing or free-running trigger pulses in a linear accelerator |
Тип |
note |
DOI |
10.1088/0031-9155/17/6/012 |
Electronic ISSN |
1361-6560 |
Print ISSN |
0031-9155 |
Журнал |
Physics in Medicine and Biology |
Том |
17 |
Первая страница |
849 |
Последняя страница |
850 |
Аффилиация |
D O Bottrill; Inst. Radiology & Nuclear Medicine, Baghdad, Iraq |
Аффилиация |
M A Jamal; Inst. Radiology & Nuclear Medicine, Baghdad, Iraq |
Аффилиация |
K I Ahmed; Inst. Radiology & Nuclear Medicine, Baghdad, Iraq |
Выпуск |
6 |