Дата выпуска 1973-03-01
dc.description The performance of silicon dosemeters is greatly affected by the ambient temperature. The expected variation in sensitivity with temperature is considered and compared with experimental results for a number of coaxial and planar devices; certain anomalies are observed at low dose rates for load resistances less than about 100 k Omega . It is suggested that these are largely due to thermoelectric EMFS arising from small (<1 degrees C) temperature differences between different parts of the system. The choice of input impedance, the sensitivity obtainable and the effect of mains pick-up are also considered.
Формат application.pdf
Издатель Institute of Physics Publishing
Название The influence of temperature on the sensitivity of silicon surface barrier detectors used for X- and gamma-ray dosimetry
Тип paper
DOI 10.1088/0031-9155/18/2/007
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 18
Первая страница 235
Последняя страница 245
Аффилиация A A Petushkov; Royal Marsden Hospital, Sutton, UK
Аффилиация R P Parker; Royal Marsden Hospital, Sutton, UK
Выпуск 2