Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор W A Langmead
Автор B F Wall
Дата выпуска 1976-01-01
dc.description A TLD system based on lithium borate which has been used both in powder form and made up into solid discs by incorporation within PTFE is described. The predominant advantages of this material over other more widely used TLD materials are its closer approximation to soft tissue, its simple glow curve which eliminates the need for complicated annealing procedures and its low cost. The luminescence from the lithium borate is measured with a specially designed reader which uses radiofrequency heating. At present, using lithium borate powder, the system is capable of measuring doses down to 10 mrad and using lithium borate PTFE discs it can measure down to 50 mrad with adequate precision. Preliminary measurements on solid glassy discs of lithium borate suggest that this form of the material may in some respects be superior for patient dosimetry.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A TLD system based on lithium borate for the measurement of doses to patients undergoing medical irradiation
Тип paper
DOI 10.1088/0031-9155/21/1/003
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 21
Первая страница 39
Последняя страница 51
Аффилиация W A Langmead; Nat. Radiological Protection Board, Harwell, Didcot, UK
Аффилиация B F Wall; Nat. Radiological Protection Board, Harwell, Didcot, UK
Выпуск 1

Скрыть метаданые