Автор R F Johnson Jr
Автор F D Rollo
Дата выпуска 1978-07-01
dc.description Commercially available <sup>81</sup>Rb and <sup>43</sup>K were evaluated objectively, that is, in terms of their effects on instrument imaging characteristics such as relative spatial resolution, plane source sensitivity and ability to demonstrate accurately various size spherical voids of radioactivity in the image plane. The rectilinear scanner was the instrument of choice because it is more readily adapted for counting the high energy photons of <sup>81</sup>Rb and <sup>43</sup>K. The parameters evaluated for both radionuclides included various pulse height analyser energy windows and collimator types. The relative performance of the scanner for each radio-nuclide was used to determine the best energy window-collimator combination for imaging.
Формат application.pdf
Издатель Institute of Physics Publishing
Название An evaluation of <sup>81</sup>Rb and <sup>43</sup>K for imaging using the performance index technique
Тип note
DOI 10.1088/0031-9155/23/4/018
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 23
Первая страница 761
Последняя страница 767
Аффилиация R F Johnson Jr; Nuclear Medicine Div., Univ. of Texas Medical Branch, Galveston, TX, USA
Аффилиация F D Rollo; Nuclear Medicine Div., Univ. of Texas Medical Branch, Galveston, TX, USA
Выпуск 4