Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор R J Jaszczak
Автор Lee-Tzuu Chang
Автор N A Stein
Автор F E Moore
Дата выпуска 1979-11-01
dc.description A whole-body single-photon emission computed tomography system (SPECT) consisting of two large-field-of-view scintillation cameras mounted on a rotatable gantry, a minicomputer and a display station has been designed, constructed and evaluated. In its usual mode of operation, eleven contiguous transverse sections, each 12.5 or 25 mm thick, are reconstructed from projection data acquired during a single, continuous 360 degrees rotation lasting from 2 to 22 min. A generalised filtered and weighted backprojection algorithm is used to reconstruct data obtained with conventional parallel-hole collimators in the case of body scanning, or with specially designed fan beam collimators in the case of centrally positioned organs. A simple yet effective, correction is used to compensate for the effects of gamma ray attenuation within the patient.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Whole-body single-photon emission computed tomography using dual, large-field-of-view scintillation cameras
Тип paper
DOI 10.1088/0031-9155/24/6/003
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 24
Первая страница 1123
Последняя страница 1143
Аффилиация R J Jaszczak; Searle Diagnostic Inc., Des Plaines, IL, USA
Аффилиация Lee-Tzuu Chang; Searle Diagnostic Inc., Des Plaines, IL, USA
Аффилиация N A Stein; Searle Diagnostic Inc., Des Plaines, IL, USA
Аффилиация F E Moore; Searle Diagnostic Inc., Des Plaines, IL, USA
Выпуск 6

Скрыть метаданые