Автор |
I L Pykett |
Автор |
B R Rosen |
Автор |
F S Buonanno |
Автор |
T J Brady |
Дата выпуска |
1983-06-01 |
dc.description |
The diagnostic utility of proton (<sup>1</sup>H) nuclear magnetic resonance (NMR) imaging is greatly enhanced when the image is weighted with proton relaxation time information. The authors indicate factors affecting the accuracy of such measurements (especially with respect to T<sub>1</sub> determination) and, specifically, the accuracy of one approach which has been used in their laboratory is assessed. A brief discussion pertaining to the validity of data obtained from biological material is presented. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Measurement of spin-lattice relaxation times in nuclear magnetic resonance imaging |
Тип |
note |
DOI |
10.1088/0031-9155/28/6/012 |
Electronic ISSN |
1361-6560 |
Print ISSN |
0031-9155 |
Журнал |
Physics in Medicine and Biology |
Том |
28 |
Первая страница |
723 |
Последняя страница |
729 |
Аффилиация |
I L Pykett; Dept. of Radiology, Massachusetts General Hospital, Boston, MA, USA |
Аффилиация |
B R Rosen; Dept. of Radiology, Massachusetts General Hospital, Boston, MA, USA |
Аффилиация |
F S Buonanno; Dept. of Radiology, Massachusetts General Hospital, Boston, MA, USA |
Аффилиация |
T J Brady; Dept. of Radiology, Massachusetts General Hospital, Boston, MA, USA |
Выпуск |
6 |