| dc.description | The authors have demonstrated that the direct voltage measurement technique using two fixed resistors on a bridge circuit considerably simplifies calorimeter operation in radiation dose measurement. They have also shown that this method is as accurate as the conventional null balance technique. These findings are believed to be significant for the fabrication of a portable in situ calorimeter. |
| Формат | application.pdf |
| Издатель | Institute of Physics Publishing |
| Название | Calorimeter dose determinations by direct voltage measurements on a Wheatstone-type bridge circuit |
| Тип | note |
| DOI | 10.1088/0031-9155/29/7/012 |
| Electronic ISSN | 1361-6560 |
| Print ISSN | 0031-9155 |
| Журнал | Physics in Medicine and Biology |
| Том | 29 |
| Первая страница | 885 |
| Последняя страница | 889 |
| Аффилиация | H Kubo; Dept. of Radiation Medicine, Massachusetts General Hospital Cancer Center, Harvard Medical School, Boston, MA, USA |