Computer methods for high resolution relaxation time imaging
G J Sivewright; D J PRendergast; D S Hickey; G J Sivewright; Dept. of Med. Biophys., Manchester Univ., UK; D J PRendergast; Dept. of Med. Biophys., Manchester Univ., UK; D S Hickey; Dept. of Med. Biophys., Manchester Univ., UK
Журнал:
Physics in Medicine and Biology
Дата:
1989-09-01
Аннотация:
Methods for calculating multiple point T<sub>1</sub> relaxation images are described and compared. A robust line fitting method is presented and its relevance to high resolution relaxation imaging discussed. Selective removal of data points is demonstrated to reduce systematic errors in linear models. A non-linear least squares iterative method is also developed and implemented. The results include simulated data, clinical images and a phantom study. Several of the methods are substantial improvements on existing linear techniques. The increased quality and consistency of the calculated images make them particularly appropriate for automated pattern recognition.
628.8Кб