Автор |
D R Chettle |
Автор |
M C Scott |
Автор |
L J Somervaille |
Дата выпуска |
1989-09-01 |
dc.description |
The measurement of Pb in bone by X-ray fluorescence using <sup>109</sup>Cd gamma -ray to excite K-shell X-rays is considered. Improvements to the basic system performance were made by changing the design of the source collimator and by including data from the Pb K beta <sub>1,3</sub> photopeaks, having used pile-up rejection to reduce background in their region of the spectrum. The summed variances for eight sample measurements are given: two each of four plaster-of-Paris phantoms embedded in wax. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
Improvements in the precision of in vivo bone lead measurements |
Тип |
paper |
DOI |
10.1088/0031-9155/34/9/014 |
Electronic ISSN |
1361-6560 |
Print ISSN |
0031-9155 |
Журнал |
Physics in Medicine and Biology |
Том |
34 |
Первая страница |
1295 |
Последняя страница |
1300 |
Аффилиация |
D R Chettle; Sch. of Phys. & Space Res., Birmingham Univ., UK |
Аффилиация |
M C Scott; Sch. of Phys. & Space Res., Birmingham Univ., UK |
Аффилиация |
L J Somervaille; Sch. of Phys. & Space Res., Birmingham Univ., UK |
Выпуск |
9 |