Автор M L McKelvie
Автор S B Palmer
Дата выпуска 1991-10-01
dc.description The Biot theory is used in an attempt to explain the frequency dependence of ultrasonic attenuation, f( alpha ), in cancellous bone. Measurements of f( alpha ) in samples of cancellous bone are compared with values calculated using parameters established from bone samples by statistical histomorphometry. Ultrasonic attenuation in cancellous bone correlates with trabecular bone volume but the Biot theory, although producing qualitative agreement produces quantitative results which are significantly deviant using parameters presently available. At the present time, too many of the Biot parameters are insufficiently defined for cancellous bone, to allow a complete test of the model.
Формат application.pdf
Издатель Institute of Physics Publishing
Название The interaction of ultrasound with cancellous bone
Тип paper
DOI 10.1088/0031-9155/36/10/003
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 36
Первая страница 1331
Последняя страница 1340
Аффилиация M L McKelvie; OTIB Autom. Technol., Cambridge, UK
Аффилиация S B Palmer; OTIB Autom. Technol., Cambridge, UK
Выпуск 10