Cone-beam CT for radiotherapy applications
P S Cho; R H Johnson; T W Griffin; P S Cho; Sch. of Med., Washington Univ., Seattle, WA, USA; R H Johnson; Sch. of Med., Washington Univ., Seattle, WA, USA; T W Griffin; Sch. of Med., Washington Univ., Seattle, WA, USA
Журнал:
Physics in Medicine and Biology
Дата:
1995-11-01
Аннотация:
Clinical implementation of cone-beam tomography has been hampered by the lack of two-dimensional electronic X-ray detectors that can encompass the full width of the body. We encountered the undersized detector problem in our development of a cone-beam CT system for radiotherapy applications. In order to mitigate the problem, we developed an algorithm which permits an increased reconstruction volume to be imaged using a detector of a given size. We describe the algorithm and report on its implementation using a radiotherapy simulator configured with a digital fluorography unit.