Дата выпуска 1997-02-01
dc.description The use of the isovolume manoeuvre method as a calibration technique for respiratory monitoring instrumentation that detects the movement of the ribcage and the abdominal wall is analysed based on a model of two connected cylinders whose radii and heights may vary, and evidence is presented which suggests that this calibration method is robust in most circumstances. Some possible functional forms relating the variations in cylinder radius and height are examined, and methods for obtaining calibration constants based on these functional forms, purely from measurements of variations in the cylinder radius, are presented.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A theoretical study of the robustness of the isovolume calibration method for a two-compartment model of breathing, based on an analysis of the connected cylinders model
Тип paper
DOI 10.1088/0031-9155/42/2/002
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 42
Первая страница 283
Последняя страница 291
Аффилиация A T Augousti; School of Applied Physics, Kingston University, Penrhyn Rd, Kingston, Surrey KT1 2EE, UK
Выпуск 2