Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор D G Lewis
Автор S S A Natto
Автор S J S Ryde
Автор C J Evans
Дата выпуска 1997-04-01
dc.description The Monte Carlo computer code MCNP has been used to design a moderated neutron source for in vivo neutron activation analysis of aluminium (Al) in the bones of the hand. The clinical motivation is the need to monitor Al body burden in subjects with renal dysfunction, at risk of Al toxicity. The design involves the source positioned on the central axis at one end of a cylindrical deuterium oxide moderator. The moderator is surrounded by a graphite reflector, with the hand inserted at the end of the moderator opposing the source. For a 1 mg source, 15 cm long radius moderator and 20 cm thick reflector, the estimated minimum detection limit is 0.5 mg Al for a 20 min irradiation, with an equivalent dose of 16.5 mSv to the hand. Increasing the moderator length and/or introducing a fast neutron filter (for example silicon) further reduces interference from fast-neutron-induced reactions on phosphorus in bone, at the expense of decreased fluence of the thermal neutrons which activate Al. Increased source strengths may be necessary to compensate for this decreased thermal fluence, to allow measurements to be made within an acceptable time limit for the comfort of the patient.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Monte Carlo design study of a moderated source for in vivo neutron activation analysis of aluminium
Тип paper
DOI 10.1088/0031-9155/42/4/002
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 42
Первая страница 625
Последняя страница 636
Выпуск 4

Скрыть метаданые