Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор August J DeClemente
Автор Ian Paddick
Дата выпуска 1997-06-01
dc.description This report discusses the calculation of x-ray doses in three dimensions using a treatment planning database which was measured in two dimensions only. It concerns the common assumption that wedged field profiles in the non-wedged direction are similar to open-field profiles for the same field size and depth. It shows the extent to which this assumption can lead to errors in wedge dose calculation for both solid and dynamically wedged fields on Varian linear accelerators at 6 MV. Finally it shows that this assumption tends to produce more accurate results when used to calculate doses for dynamically wedged fields and why, even in the wedged direction, some of the simpler treatment planning algorithms are more suitable for dynamic wedges than they are for solid wedges.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Errors in three-dimensional doses calculated from a two-dimensional database - case report: wedged fields at 6 MV
Тип note
DOI 10.1088/0031-9155/42/6/015
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 42
Первая страница 1197
Последняя страница 1202
Аффилиация August J DeClemente; Medical Physics Department, Cromwell Hospital, Cromwell Road, London SW5 0TU, UK
Аффилиация Ian Paddick; Medical Physics Department, Cromwell Hospital, Cromwell Road, London SW5 0TU, UK
Выпуск 6

Скрыть метаданые