dc.description In ionization chamber dosimetry close to brachytherapy sources, the non-uniform photon fluence in the vicinity of the air cavity of the ionization chamber must be corrected for. In this paper, experimentally determined relative non-uniformity correction factors are determined for a PTW23331 chamber and a number of Farmer-type chambers. It is concluded that the `anisotropic' theory of Bielajew agrees better with the experimental results than the `isotropic' theory of Kondo and Randolph. The experiments show that neither the material choice nor the radius of the central electrode have any significant effect on the non-uniformity correction factors. Experiments on the wall-material dependence in the non-uniformity correction factor, which is predicted by the anisotropic theory, were inconclusive. Parameters for the Farmer-type chambers for calculation of theoretical non-uniformity correction factors are given. Data to derive anisotropic fluence non-uniformity correction factors are given in tabular form. An error in the Kondo and Randolph equations is reported and the corrected equations presented.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Fluence non-uniformity effects in air kerma determination around brachytherapy sources
Тип paper
DOI 10.1088/0031-9155/42/7/006
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 42
Первая страница 1301
Последняя страница 1318
Выпуск 7

Скрыть метаданые