Improvements in the design of linear diffusers for photodynamic therapy
L H P Murrer; J P A Marijnissen; W M Star; L H P Murrer; Department of Clinical Physics, Dr Daniel Den Hoed Cancer Centre, Erasmus University Hospital, PO Box 5201, 3008 AE Rotterdam, The Netherlands; J P A Marijnissen; Department of Clinical Physics, Dr Daniel Den Hoed Cancer Centre, Erasmus University Hospital, PO Box 5201, 3008 AE Rotterdam, The Netherlands; W M Star; Department of Clinical Physics, Dr Daniel Den Hoed Cancer Centre, Erasmus University Hospital, PO Box 5201, 3008 AE Rotterdam, The Netherlands
Журнал:
Physics in Medicine and Biology
Дата:
1997-07-01
Аннотация:
The angular radiance distribution of several linear diffusers used for photodynamic therapy (PDT) was measured. The forward scattering found previously was not observed for these designs. The improved isotropy leads to a better agreement between intended treatment site and actual maximum of the fluence rate profile when the linear diffuser is used in a hollow, cylindrical organ.