dc.description This paper evaluates the characteristics of ionization chambers for the measurement of absorbed dose to water for medium-energy x-rays. The values of the chamber correction factor, , used in the IPEMB code of practice for the UK secondary standard (NE2561/NE2611) ionization chamber are derived and their constituent factors examined. The comparison of the chambers' responses in air revealed that of the chambers tested only the NE2561, NE2571 and NE2505 exhibit a flat (within 5%) energy response in air. Under no circumstances should the NACP, Sanders electron chamber, or any chamber that has a wall made of high atomic number material, be used for medium-energy x-ray dosimetry. The measurements in water reveal that a chamber that has a substantial housing, such as the PTW Grenz chamber, should not be used to measure absorbed dose to water in this energy range. The value of for an NE2561 chamber was determined by measuring the absorbed dose to water and comparing it with that for an NE2571 chamber, for which data have been published. The chamber correction factor varies from to for x-ray beams with HVL between 0.15 and 4 mm Cu. The values agree with that for an NE2571 chamber within the experimental uncertainty. The corrections due to the stem, waterproof sleeve and replacement of the phantom material by the chamber for an NE2561 chamber are described.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Investigation of the chamber correction factor (k<sub>ch</sub>) for the UK secondary standard ionization chamber (NE2561/NE2611) using medium-energy x-rays
Тип paper
DOI 10.1088/0031-9155/43/11/003
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 43
Первая страница 3195
Последняя страница 3206
Аффилиация K E Rosser; National Physical Laboratory, Teddington, Middlesex TW11 OLW, UK
Выпуск 11

Скрыть метаданые