dc.description Measuring portal dose with an electronic portal imaging device (EPID) in external beam radiotherapy can be used to perform routine dosimetric quality control checks on linear accelerators and to verify treatments ( in vivo dosimetry). An accurate method to measure portal dose images (PDIs) with a commercially available fluoroscopic EPID has been developed. The method accounts for (i) the optical `cross talk' within the EPID structure, (ii) the spatially non-uniform EPID response and (iii) the nonlinearity of the EPID response. The method is based on a deconvolution algorithm. Measurement of the required input data is straightforward. The observed nonlinearity of the EPID response was largely due to the somewhat outdated EPID electronics. Nonlinearity corrections for more modern systems are expected to be smaller. The accuracy of the method was assessed by comparing PDIs measured with the EPID with PDIs measured with a scanning ionization chamber in a miniphantom, located at the same position as the fluorescent screen. For irradiations in open, wedged and intensity modulated 25 MV photon beams (produced with dynamic multileaf collimation) EPID and ionization chamber measurements agreed to within 1% (1 SD).
Формат application.pdf
Издатель Institute of Physics Publishing
Название Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation
Тип paper
DOI 10.1088/0031-9155/43/8/004
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 43
Первая страница 2047
Последняя страница 2060
Аффилиация K L Pasma; Department of Clinical Physics, Daniel den Hoed Cancer Center/University Hospital Rotterdam, Groene Hilledijk 301, 3075 EA Rotterdam, The Netherlands
Аффилиация M Kroonwijk; Department of Clinical Physics, Daniel den Hoed Cancer Center/University Hospital Rotterdam, Groene Hilledijk 301, 3075 EA Rotterdam, The Netherlands
Аффилиация J C J de Boer; Department of Clinical Physics, Daniel den Hoed Cancer Center/University Hospital Rotterdam, Groene Hilledijk 301, 3075 EA Rotterdam, The Netherlands
Аффилиация A G Visser; Department of Clinical Physics, Daniel den Hoed Cancer Center/University Hospital Rotterdam, Groene Hilledijk 301, 3075 EA Rotterdam, The Netherlands
Аффилиация B J M Heijmen; Department of Clinical Physics, Daniel den Hoed Cancer Center/University Hospital Rotterdam, Groene Hilledijk 301, 3075 EA Rotterdam, The Netherlands
Выпуск 8

Скрыть метаданые