Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор Patricia B Scott
Автор J R Greening
Дата выпуска 1963-04-01
dc.description The usual method of determining saturation currents in free-air chambers is to plot the reciprocal of the current, j, against the reciprocal of the collecting electric field, E, and extrapolate linearly to 1/E=0. It is shown that this method is valid only if the recombination taking place is initial recombination within ion clusters. Experimental and theoretical evidence is produced to show that in such chambers, at dose rates of 1 r/min or more, it is general (volume) rather than initial recombination which is taking place. Theory shows that under these circumstances one should plot 1/j against 1/E<sup>2</sup>, or more accurately against j/E<sup>2</sup>, and extrapolate to 1/E<sup>2</sup> or j/E<sup>2</sup>=0.
Формат application.pdf
Издатель Institute of Physics Publishing
Название The Determination of Saturation Currents in Free-air Ionization Chambers by Extrapolation Methods
Тип paper
DOI 10.1088/0031-9155/8/1/304
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 8
Первая страница 51
Последняя страница 57
Аффилиация Patricia B Scott; Department of Medical Physics, University of Edinburgh
Аффилиация J R Greening; Department of Medical Physics, University of Edinburgh
Выпуск 1

Скрыть метаданые