Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор L A W Kemp
Автор Barry Barber
Дата выпуска 1963-06-01
dc.description A discussion of the merits of the guarded-field type of ionization chamber as compared with the conventional type of thimble chamber is illustrated, with outlines of the electrostatic field distributions. Details of the mode of construction of a practical and robust guarded-field chamber are given. With the exposure dose rates (of the order of 100 r per minute) for which the chamber was designed it becomes feasible and advantageous to use the chamber in conjunction with a high-sensitivity moving-coil galvanometer instead of a Townsend Balance. A suitable galvanometer is described, together with a calibrating network which enables the ionization current to be determined by a null-type matching procedure.The results of Boag's theory of general recombination in a parallel plate type of chamber are summarized and it is shown that, as predicted by the theory, a plot of the current, j, against the square of the ratio of the current to the polarizing voltage (j/V)<sup>2</sup>, yields a straight line which can be extrapolated to obtain the saturation current. The result of an absolute measurement of the exposure dose rate of The London Hospital cobalt unit is compared with the output as determined by a Victrometer dosemeter calibrated at the National Physical Laboratory.
Формат application.pdf
Издатель Institute of Physics Publishing
Название The Construction and Use of a Guarded-Field Cavity Ionization Chamber for the Measurement of Supervoltage Radiation
Тип paper
DOI 10.1088/0031-9155/8/2/302
Electronic ISSN 1361-6560
Print ISSN 0031-9155
Журнал Physics in Medicine and Biology
Том 8
Первая страница 149
Последняя страница 160
Аффилиация L A W Kemp; Physics Department, The London Hospital Research Laboratories, Ashfield Street, London, E.1
Аффилиация Barry Barber; Physics Department, The London Hospital Research Laboratories, Ashfield Street, London, E.1
Выпуск 2

Скрыть метаданые