Автор H T Jessop
Дата выпуска 1953-05-01
dc.description An investigation into the errors in measurements by the Tardy and Sénarmont methods shows: (i) that the errors due to inaccurate quarter-wave plates are less in the Sénarmont method than in the Tardy, (ii) that errors due to inaccurate setting of the polaroids are of the same order in both methods, and (iii) that in general the error in the final setting of the analyser is the only one which need be considered. A full account is given of the less well-known Sénarmont method.
Формат application.pdf
Издатель Institute of Physics Publishing
Название On the Tardy and Sénarmont methods of measuring fractional relative retardations
Тип paper
DOI 10.1088/0508-3443/4/5/303
Print ISSN 0508-3443
Журнал British Journal of Applied Physics
Том 4
Первая страница 138
Последняя страница 141
Аффилиация H T Jessop; University College, London
Выпуск 5

Скрыть метаданые