Дата выпуска 1934-10-01
dc.description In mounting a moving coil to give a variation of 1 per cent. or so on a mutual inductance, it is necessary to determine at least two quantities - the length of the arm carrying the moving coil and the separation of its axis of rotation from the axis of the fixed coil. A simple method is described for the simultaneous determination of these quantities, only one preliminary set of experimental results being necessary. A simple reciprocal property is pointed out which leads to an improvement in mechanical design.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A simple method of selecting an inductometer scale
Тип paper
DOI 10.1088/0950-7671/11/10/304
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 11
Первая страница 319
Последняя страница 320
Аффилиация N F Astbury; Electricity Department, National Physical Laboratory, Teddington, Middlesex