Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор R S Higginbotham
Автор J J Benbow
Дата выпуска 1952-07-01
dc.description Two modifications of the cone and plate viscometer are described. The first, for measurements at low shearing stresses (10 to 100 dyn/cm<sup>2</sup>), employs a phosphor-bronze strip to apply torque to the cone. The second, a direct reading instrument for measurements at a relatively high stress (4 500 dyn/cm<sup>2</sup>), uses a small d.c. motor running at constant torque to drive the cone. The range of the first instrument is 1 to 100 poises at 10 dyn/cm<sup>2</sup>, and 10 to 100 poises at 100 dyn/cm<sup>2</sup>, that of the second is 0.5 to 30 poises. Their accuracy varies according to the viscosity and anomaly of the fluid under examination, but the average is about ± 5%, which is quite adequate for many purposes. The examination of a sample with either instrument, including filling and cleaning, can be completed within two minutes.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Two viscometers for rapid measurements at definite shearing stresses
Тип paper
DOI 10.1088/0950-7671/29/7/307
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 29
Первая страница 221
Последняя страница 224
Аффилиация R S Higginbotham; The British Cotton Industry Research Association. Shirley Institute, Manchester, 20
Аффилиация J J Benbow; The British Cotton Industry Research Association. Shirley Institute, Manchester, 20
Выпуск 7

Скрыть метаданые