Автор |
B N Audric |
Автор |
J V P Long |
Дата выпуска |
1953-12-01 |
dc.description |
Experiments with scintillation counters using p-terphenyl in benzene are described. By the use of a combination of steel screening, coincidence counting and pulse height selection the total background is reduced to 30 c.p.m. at room temperature for a counting volume of 20 ml. The <sup>14</sup>C detection efficiency under these conditions is approximately 40%. Light interaction between the photomultipliers used is almost eliminated by the use of a suitable filter. |
Формат |
application.pdf |
Издатель |
Institute of Physics Publishing |
Название |
The background and <sup>14</sup>C detection efficiency of a liquid scintillation counter |
Тип |
paper |
DOI |
10.1088/0950-7671/30/12/309 |
Print ISSN |
0950-7671 |
Журнал |
Journal of Scientific Instruments |
Том |
30 |
Первая страница |
467 |
Последняя страница |
469 |
Аффилиация |
B N Audric; Chemical Research Laboratory, Teddington, Middlesex |
Аффилиация |
J V P Long; Chemical Research Laboratory, Teddington, Middlesex |
Выпуск |
12 |