Автор J V P Long
Дата выпуска 1958-09-01
dc.description A point focus X-ray tube has been used as a source for microanalysis by the X-ray absorption method. Transmission measurements on selected specimen areas of 2-10 μ diameter are possible, with wavelengths chosen by a crystal and a proportional counter. Comparison of the transmitted intensities at different wavelengths gives quantitative information on the mass thickness of elements present in the area examined. The ultimate mass sensitivity varies with the position of the element in the periodic table but is usually in the range 10<sup>-10</sup>-10<sup>-12</sup> g for elements of Z > 15. The application of the method to the determination of calcium is described.
Формат application.pdf
Издатель Institute of Physics Publishing
Название X-ray absorption microanalysis with fine-focus tubes
Тип paper
DOI 10.1088/0950-7671/35/9/303
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 35
Первая страница 323
Последняя страница 329
Аффилиация J V P Long; Cavendish Laboratory, University of Cambridge