dc.description Electrons with a mean energy of a few electron volts emerging from a source a few millimetres in diameter are accelerated on to a positively biased plastic scintillator. The light generated in the scintillator is guided by a Perspex light-pipe to a photomultiplier. If the scintillator bias is greater than 10 kV, no noise is introduced by the scintillator-light-pipe-photomultiplier system. If the original electron signal is modulated, the maximum modulation frequency transmitted by the system is greater than 10 Mc/s. The minimum detectable current is limited by the photomultiplier dark current, and may be less than 10<sup>-15</sup> A with a suitably chosen tube.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Wide-band detector for micro-microampere low-energy electron currents
Тип paper
DOI 10.1088/0950-7671/37/7/307
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 37
Первая страница 246
Последняя страница 248
Аффилиация T E Everhart; Department of Engineering, University of Cambridge