Мобильная версия

Доступно журналов:

3 288

Доступно статей:

3 891 637

 

Скрыть метаданые

Автор J C van der Breggen
Автор H J Wouterlood
Дата выпуска 1960-08-01
dc.description A transducer has been developed to detect changes in position of a balance beam. Its operation is based upon a differential capacitor. Variations in capacitance are made to produce directly a d.c. voltage corresponding in sign to the sense of deflexion of the balance beam. The construction of the transducer is described, and its application in automatic apparatus for thermogravimetric analysis is illustrated by results obtained on the dehydration of copper sulphate and the decomposition of quinoline phosphomolybdate.The sensitivity of the instrument can be varied in nine steps to give full-scale deflexions from 10 to 5000 mg. An accuracy of about 1.5 mg or 0.75% of full scale was obtained for the 200 mg range.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Novel displacement detector and its application in a recording thermobalance
Тип paper
DOI 10.1088/0950-7671/37/8/315
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 37
Первая страница 297
Последняя страница 300
Аффилиация J C van der Breggen; African Explosives and Chemical Industries Ltd., Northrand, Transvaal, South Africa
Аффилиация H J Wouterlood; African Explosives and Chemical Industries Ltd., Northrand, Transvaal, South Africa
Выпуск 8