Metal cell for use in vapour pressure measurements by the effusion method
S Blairs; R A J Shelton; R Unsworth; S Blairs; Dept. of Metallurgy, The University, Manchester 13; R A J Shelton; Dept. of Metallurgy, The University, Manchester 13; R Unsworth; Dept. of Metallurgy, The University, Manchester 13
Журнал:
Journal of Scientific Instruments
Дата:
1961-12-01
Аннотация:
Full constructional details are given for an effusion cell in Armco iron, which in conjunction with a calibrated glass spring and cathetometer has been used to measure the vapour pressure over solid ferric chloride in the temperature range 100-200° C. The chief advantage claimed for the cell is that the effusion orifices, in interchangeable push fit end-caps, are of the form of true circular knife edged holes with easily determined dimensions.