dc.description It has been shown that the conventional practice of designing a compensated hot wire amplifier with a fixed ceiling to floor ratio results in considerable and unnecessary increase in noise level at compensation settings other than optimum (which is at the maximum compensation at the highest frequency of interest). The optimum ceiling to floor ratio has been estimated to be between 1.5-2.0 ω<sub>max</sub>M. Application of the above considerations to an amplifier in which the ceiling to floor ratio is optimized at each compensation setting (for a given amplifier band-width), shows the usefulness of the method in improving the signal to noise ratio.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Method for obtaining minimum noise in a general purpose hot wire amplifier
Тип paper
DOI 10.1088/0950-7671/38/12/318
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 38
Первая страница 491
Последняя страница 492
Аффилиация G N V Rao; Department of Aeronautical Engineering, Indian Institute of Science, Bangalore 12, India
Выпуск 12