dc.description A new millimetre wave magnetic resonance spectrometer, using broad-band components, is described. A `magic tee' provides both a balanced harmonic generator and a detector of the harmonic signal which has passed through the sample in the spectrometer. Studies have been made of a simple reflection system with no resonant cavity at about 4.7 mm wave-length, using standard 8 mm components. With a power of about 30 MW incident on the harmonic generator crystals, a time constant of 1 sec and a magnetic field modulation frequency of 630 c/s, approximately 9 × 10<sup>15</sup> paramagnetic spins of 1 gauss line-width can be detected. The device is primarily intended for antiferromagnetic resonance studies and is in principle capable of operating at much shorter millimetre wavelengths.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A novel millimetre wave magnetic resonance spectrometer
Тип paper
DOI 10.1088/0950-7671/40/12/318
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 40
Первая страница 586
Последняя страница 588
Аффилиация P R Elliston; Physics Department, Monash University, Clayton, Victoria, Australia
Аффилиация G J Troup; Physics Department, Monash University, Clayton, Victoria, Australia
Аффилиация D R Hutton; Physics Department, Monash University, Clayton, Victoria, Australia
Выпуск 12

Скрыть метаданые