| Автор | Joan H Dunmore |
| Автор | R J Hamilton |
| Автор | D S G Smith |
| Дата выпуска | 1964-11-01 |
| dc.description | The construction and performance of an instrument which is portable, self-powered, intrinsically safe in mine atmospheres and designed to be left to operate unattended during a working shift are described.Following its aerodynamic selection by passage through a `horizontal elutriator', the `respirable' dust is collected on a glass fibre paper filter through which air is drawn by a reciprocating diaphragm pump driven by a governed 6 v motor. Air flow rate (2.5 l. min<sup>-1</sup>) is largely independent of build-up in filter resistance, and the glass fibre paper is little affected by ambient moisture. Simple estimation of coal/non-coal content of the sample is given by incineration of the dust on the filter. |
| Формат | application.pdf |
| Издатель | Institute of Physics Publishing |
| Название | An instrument for the sampling of respirable dust for subsequent gravimetric assessment |
| Тип | paper |
| DOI | 10.1088/0950-7671/41/11/304 |
| Print ISSN | 0950-7671 |
| Журнал | Journal of Scientific Instruments |
| Том | 41 |
| Первая страница | 669 |
| Последняя страница | 672 |
| Аффилиация | Joan H Dunmore; Mining Research Establishment, Isleworth, Middlesex |
| Аффилиация | R J Hamilton; Mining Research Establishment, Isleworth, Middlesex |
| Аффилиация | D S G Smith; Mining Research Establishment, Isleworth, Middlesex |
| Выпуск | 11 |