Дата выпуска 1964-04-01
dc.description With the instrument described, the dew point of gas flowing in a tube is obtained by observing the temperature at which condensation is formed on a mirror surface that is heated and cooled electrically. Dew points 20 degc below the gas temperature and in the range 50 to -5° C have been measured in gas flows of 40 1. min<sup>-1</sup>.
Формат application.pdf
Издатель Institute of Physics Publishing
Название A dew point meter based on thermoelectric cooling
Тип note
DOI 10.1088/0950-7671/41/4/416
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 41
Первая страница 241
Последняя страница 241
Аффилиация O G Griffin; Safety in Mines Research Establishment, Ministry of Power, Sheffield 3
Аффилиация C M Stringfield; Safety in Mines Research Establishment, Ministry of Power, Sheffield 3
Выпуск 4

Скрыть метаданые