Дата выпуска 1965-09-01
dc.description The conversion of a 40 Mc/s Varian nuclear magnetic resonance spectrometer to 80 Mc/s operation is described. The increase in maximum field strength from 12 to 22.5 kG was achieved by the use of conical Permendur pole caps. The factors affecting pole-cap geometry are discussed. Sufficient field homogeneity for the production of high resolution 80 Mc/s hydrogen nuclear magnetic resonance spectra was achieved with Golay electric shim coils. Further stability of the magnetic field, enabling spectra to be recorded on precalibrated chart paper, was achieved by a double audio-frequency modulation technique. Notes are also given on the associated changes required to the electronic circuits.
Формат application.pdf
Издатель Institute of Physics Publishing
Название Conversion of a Varian 40 Mc/s high resolution nuclear magnetic resonance spectrometer to 80 Mc/s operation
Тип paper
DOI 10.1088/0950-7671/42/9/301
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 42
Первая страница 669
Последняя страница 674
Аффилиация S A Knight; The British Petroleum Co. Ltd., Sunbury-on-Thames, Middlesex
Аффилиация R L Erskine; The British Petroleum Co. Ltd., Sunbury-on-Thames, Middlesex
Выпуск 9