Автор G Szabó Miszenti
Дата выпуска 1966-11-01
dc.description A particularly simple application of an original oscillation control system to a low-frequency torsion pendulum (equipped) for non-automatic discontinuous measurement of internal friction by the free decay method) has made possible the automatic continuous record of Q<sup>-1</sup> as a function of time at practically constant amplitude. Q<sup>-1</sup> can be measured up to at least 10<sup>-2</sup>. The precision is typically to better than 3 × 10<sup>-5</sup> + 0.5%. The method also allows period measurement within less than 0.1 msec at 1 c/s. Some results on the relaxation effects due to about 0.002 wt% of dissolved nitrogen in α-Fe are reported, as an example of the instrument performance. Systematic errors of thermocouple temperature measurement in non-stationary conditions are discussed.
Формат application.pdf
Издатель Institute of Physics Publishing
Название An automatic system for continuous measurement of internal friction and oscillation period at low frequency (1 c/s range)
Тип paper
DOI 10.1088/0950-7671/43/11/308
Print ISSN 0950-7671
Журнал Journal of Scientific Instruments
Том 43
Первая страница 820
Последняя страница 824
Аффилиация G Szabó Miszenti; Sorin, Centro Richerche Nucleari, Saluggia, Italy
Выпуск 11